The IIO standard units are measured in kPa while the driver is using hPa. Apart from checking the userspace value itself, it is mentioned also in the Bosch API [1] that the pressure value is in Pascal. [1]: https://github.com/boschsensortec/BME68x_SensorAPI/blob/v4.4.8/bme68x_defs.h#L742 Fixes: 1b3bd8592780 ("iio: chemical: Add support for Bosch BME680 sensor") Signed-off-by: Vasileios Amoiridis <vassilisamir@xxxxxxxxx> --- drivers/iio/chemical/bme680_core.c | 2 +- 1 file changed, 1 insertion(+), 1 deletion(-) diff --git a/drivers/iio/chemical/bme680_core.c b/drivers/iio/chemical/bme680_core.c index ef5e0e46fd34..2c40c13fe97a 100644 --- a/drivers/iio/chemical/bme680_core.c +++ b/drivers/iio/chemical/bme680_core.c @@ -678,7 +678,7 @@ static int bme680_read_press(struct bme680_data *data, } *val = bme680_compensate_press(data, adc_press); - *val2 = 100; + *val2 = 1000; return IIO_VAL_FRACTIONAL; } base-commit: b3019fcdeb286b2cfe45e44bccb44dbcd8ff66dd -- 2.25.1