Add honeywell,pressure-triplet property that automatically initializes pmin-pascal, pmax-pascal so that the user is not required to look-up the chip in the datasheet and convert various units to pascals himself. Signed-off-by: Petre Rodan <petre.rodan@xxxxxxxxxxxxxxx> Signed-off-by: Andreas Klinger <ak@xxxxxxxxxxxxx> --- drivers/iio/pressure/mprls0025pa.c | 102 +++++++++++++++++++++++++++-- 1 file changed, 95 insertions(+), 7 deletions(-) diff --git a/drivers/iio/pressure/mprls0025pa.c b/drivers/iio/pressure/mprls0025pa.c index cb5d6c0cca7e..90aed290b6c2 100644 --- a/drivers/iio/pressure/mprls0025pa.c +++ b/drivers/iio/pressure/mprls0025pa.c @@ -81,6 +81,78 @@ static const struct mpr_func_spec mpr_func_spec[] = { [MPR_FUNCTION_C] = { .output_min = 3355443, .output_max = 13421773 }, }; +enum mpr_variants { + MPR0001BA = 0x00, MPR01_6BA = 0x01, MPR02_5BA = 0x02, MPR0060MG = 0x03, + MPR0100MG = 0x04, MPR0160MG = 0x05, MPR0250MG = 0x06, MPR0400MG = 0x07, + MPR0600MG = 0x08, MPR0001BG = 0x09, MPR01_6BG = 0x0a, MPR02_5BG = 0x0b, + MPR0100KA = 0x0c, MPR0160KA = 0x0d, MPR0250KA = 0x0e, MPR0006KG = 0x0f, + MPR0010KG = 0x10, MPR0016KG = 0x11, MPR0025KG = 0x12, MPR0040KG = 0x13, + MPR0060KG = 0x14, MPR0100KG = 0x15, MPR0160KG = 0x16, MPR0250KG = 0x17, + MPR0015PA = 0x18, MPR0025PA = 0x19, MPR0030PA = 0x1a, MPR0001PG = 0x1b, + MPR0005PG = 0x1c, MPR0015PG = 0x1d, MPR0030PG = 0x1e, MPR0300YG = 0x1f, + MPR_VARIANTS_MAX +}; + +static const char * const mpr_triplet_variants[MPR_VARIANTS_MAX] = { + [MPR0001BA] = "0001BA", [MPR01_6BA] = "01.6BA", [MPR02_5BA] = "02.5BA", + [MPR0060MG] = "0060MG", [MPR0100MG] = "0100MG", [MPR0160MG] = "0160MG", + [MPR0250MG] = "0250MG", [MPR0400MG] = "0400MG", [MPR0600MG] = "0600MG", + [MPR0001BG] = "0001BG", [MPR01_6BG] = "01.6BG", [MPR02_5BG] = "02.5BG", + [MPR0100KA] = "0100KA", [MPR0160KA] = "0160KA", [MPR0250KA] = "0250KA", + [MPR0006KG] = "0006KG", [MPR0010KG] = "0010KG", [MPR0016KG] = "0016KG", + [MPR0025KG] = "0025KG", [MPR0040KG] = "0040KG", [MPR0060KG] = "0060KG", + [MPR0100KG] = "0100KG", [MPR0160KG] = "0160KG", [MPR0250KG] = "0250KG", + [MPR0015PA] = "0015PA", [MPR0025PA] = "0025PA", [MPR0030PA] = "0030PA", + [MPR0001PG] = "0001PG", [MPR0005PG] = "0005PG", [MPR0015PG] = "0015PG", + [MPR0030PG] = "0030PG", [MPR0300YG] = "0300YG" +}; + +/** + * struct mpr_range_config - list of pressure ranges based on nomenclature + * @pmin: lowest pressure that can be measured + * @pmax: highest pressure that can be measured + */ +struct mpr_range_config { + const s32 pmin; + const s32 pmax; +}; + +/* All min max limits have been converted to pascals */ +static const struct mpr_range_config mpr_range_config[MPR_VARIANTS_MAX] = { + [MPR0001BA] = { .pmin = 0, .pmax = 100000 }, + [MPR01_6BA] = { .pmin = 0, .pmax = 160000 }, + [MPR02_5BA] = { .pmin = 0, .pmax = 250000 }, + [MPR0060MG] = { .pmin = 0, .pmax = 6000 }, + [MPR0100MG] = { .pmin = 0, .pmax = 10000 }, + [MPR0160MG] = { .pmin = 0, .pmax = 16000 }, + [MPR0250MG] = { .pmin = 0, .pmax = 25000 }, + [MPR0400MG] = { .pmin = 0, .pmax = 40000 }, + [MPR0600MG] = { .pmin = 0, .pmax = 60000 }, + [MPR0001BG] = { .pmin = 0, .pmax = 100000 }, + [MPR01_6BG] = { .pmin = 0, .pmax = 160000 }, + [MPR02_5BG] = { .pmin = 0, .pmax = 250000 }, + [MPR0100KA] = { .pmin = 0, .pmax = 100000 }, + [MPR0160KA] = { .pmin = 0, .pmax = 160000 }, + [MPR0250KA] = { .pmin = 0, .pmax = 250000 }, + [MPR0006KG] = { .pmin = 0, .pmax = 6000 }, + [MPR0010KG] = { .pmin = 0, .pmax = 10000 }, + [MPR0016KG] = { .pmin = 0, .pmax = 16000 }, + [MPR0025KG] = { .pmin = 0, .pmax = 25000 }, + [MPR0040KG] = { .pmin = 0, .pmax = 40000 }, + [MPR0060KG] = { .pmin = 0, .pmax = 60000 }, + [MPR0100KG] = { .pmin = 0, .pmax = 100000 }, + [MPR0160KG] = { .pmin = 0, .pmax = 160000 }, + [MPR0250KG] = { .pmin = 0, .pmax = 250000 }, + [MPR0015PA] = { .pmin = 0, .pmax = 103421 }, + [MPR0025PA] = { .pmin = 0, .pmax = 172369 }, + [MPR0030PA] = { .pmin = 0, .pmax = 206843 }, + [MPR0001PG] = { .pmin = 0, .pmax = 6895 }, + [MPR0005PG] = { .pmin = 0, .pmax = 34474 }, + [MPR0015PG] = { .pmin = 0, .pmax = 103421 }, + [MPR0030PG] = { .pmin = 0, .pmax = 206843 }, + [MPR0300YG] = { .pmin = 0, .pmax = 39997 } +}; + static const struct iio_chan_spec mpr_channels[] = { { .type = IIO_PRESSURE, @@ -258,6 +330,7 @@ int mpr_common_probe(struct device *dev, const struct mpr_ops *ops, int irq) int ret; struct mpr_data *data; struct iio_dev *indio_dev; + const char *triplet; s64 scale, offset; u32 func; @@ -299,17 +372,32 @@ int mpr_common_probe(struct device *dev, const struct mpr_ops *ops, int irq) "honeywell,transfer-function %d invalid\n", data->function); - ret = device_property_read_u32(dev, "honeywell,pmin-pascal", + ret = device_property_read_string(dev, "honeywell,pressure-triplet", + &triplet); + if (ret) { + ret = device_property_read_u32(dev, "honeywell,pmin-pascal", &data->pmin); - if (ret) - return dev_err_probe(dev, ret, + if (ret) + return dev_err_probe(dev, ret, "honeywell,pmin-pascal could not be read\n"); - ret = device_property_read_u32(dev, "honeywell,pmax-pascal", - &data->pmax); - if (ret) - return dev_err_probe(dev, ret, + ret = device_property_read_u32(dev, "honeywell,pmax-pascal", + &data->pmax); + if (ret) + return dev_err_probe(dev, ret, "honeywell,pmax-pascal could not be read\n"); + } else { + ret = device_property_match_property_string(dev, + "honeywell,pressure-triplet", + mpr_triplet_variants, + MPR_VARIANTS_MAX); + if (ret < 0) + return dev_err_probe(dev, -EINVAL, + "honeywell,pressure-triplet is invalid\n"); + + data->pmin = mpr_range_config[ret].pmin; + data->pmax = mpr_range_config[ret].pmax; + } if (data->pmin >= data->pmax) return dev_err_probe(dev, -EINVAL, -- 2.41.0