You are subscribed to Recalls, Market Withdrawals and Safety Alerts for U.S. Food & Drug Administration (FDA). This information has recently been updated and is now available.
08/24/2017 08:58 PM EDT
Cook Medical initiated a correction to the Instructions For Use (IFU) for the Zenith Alpha™ Thoracic Endovascular Graft. This correction removed the indication for blunt thoracic aortic injury, also known as BTAI or “transection” of the aorta because Cook has received an increase in reports of graft thrombosis and occlusion with these grafts specifically in the treatment of BTAI. Cook also initiated a voluntary recall of Zenith Alpha Thoracic products in sizes of 18-22mm, including the 26-22mm tapered device. These specific sizes are primarily used for BTAI, and are no longer available for purchase.
For detailed information pertaining to this Recalls, Market Withdrawals and Safety Alerts message, please click the link at the beginning of this bulletin. |