Medical Device Safety and Recalls: B. Braun Medical Inc. Recalls Dialog+ Hemodialysis Systems Due Defective Conductivity Sensors

[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

 



Title: Medical Device Safety and Recalls: B. Braun Medical Inc. Recalls Dialog+ Hemodialysis Systems Due Defective Conductivity Sensors

B. Braun Medical Inc. is recalling the Dialog+ Hemodialysis System due to cracks in conductivity sensors that may allow air to enter into the solution (dialysis fluid or dialysate) used to help filter waste and other excess fluids in the blood. The presence of air in dialysis fluid may lead to improper blood filtration, causing serious adverse health consequences, including death.


This email was sent to list-fda@xxxxxxxxxxx using GovDelivery, on behalf of: U.S. Food & Drug Administration (FDA) · 10903 New Hampshire Ave · Silver Spring, MD 20993 · 1-888-INFO-FDA (1-888-463-6332) Powered by GovDelivery