FDA MedWatch - Dialog+ Hemodialysis Systems by B. Braun Medical: Class I Recall - Defective Conductivity Sensors

[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

 



Title: FDA MedWatch - Dialog+ Hemodialysis Systems by B. Braun Medical: Class I Recall - Defective Conductivity Sensors
MedWatch logoMedWatch - The FDA Safety Information and Adverse Event Reporting Program

Dialog+ Hemodialysis Systems by B. Braun Medical: Class I Recall - Defective Conductivity Sensors

AUDIENCE: Risk Manager, Nephrology

ISSUE: B. Braun Medical Inc. is recalling the Dialog+ Hemodialysis System due to cracks in conductivity sensors that may allow air to enter into the solution (dialysis fluid or dialysate) used to help filter waste and other excess fluids in the blood. The presence of air in dialysis fluid may lead to improper blood filtration, causing serious adverse health consequences, including death.

The recalled products include

BACKGROUND: The Dialog+ Hemodialysis System is a machine used in the treatment of chronic kidney disease for patients whose own kidneys are no longer healthy enough to filter their blood of wastes and excess fluid. These systems are used in hospitals, health centers, and in outpatient dialysis center settings.

RECOMMENDATION: On April 1, 2016, B. Braun Inc. sent an “Urgent Medical Device Correction” letter to affected customers asking them to:

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

Read the MedWatch alert, including a link to the FDA Class I Recall Notice, at:

http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm499143.htm