Medical Device Safety and Recalls: Dräger Evita V500 and Babylog VN500 Ventilators - Issue with Optional PS500 Battery Power Supply May Cause Ventilators to Shut Down Unexpectedly

[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

 



Title: Medical Device Safety and Recalls: Dräger Evita V500 and Babylog VN500 Ventilators - Issue with Optional PS500 Battery Power Supply May Cause Ventilators to Shut Down Unexpectedly

A recall has been posted regarding Dräger Evita V500 and Babylog VN500 Ventilators.  Dräger is recalling the PS500 battery power supply because a software issue causes shorter than expected battery run times.  The software issue with the power supply also prevents the appropriate alarm from sounding 5 minutes before the battery runs out of power and the device shuts down.  If the ventilator shuts down, a patient may not receive necessary oxygen. This could cause patient injury or death.


This email was sent to list-fda@xxxxxxxxxxx using GovDelivery, on behalf of: U.S. Food & Drug Administration (FDA) · 10903 New Hampshire Ave · Silver Spring, MD 20993 · 1-888-INFO-FDA (1-888-463-6332) Powered by GovDelivery