Treanda (bendamustine hydrochloride) Solution by Teva: FDA Statement - Not Compatible with Closed System Transfer Devices, Adapters, and Syringes Containing Polycarbonate or Acrylonitrile-Butadiene-Styrene

AUDIENCE: Oncology, Pharmacy, Nursing, Risk Manager

ISSUE: FDA is warning health care professionals not to use Treanda Injection (45 mg/0.5 mL or 180 mg/2 mL solution) with closed system transfer devices (CSTD), adapters, and syringes containing polycarbonate or acrylonitrile-butadiene-styrene (ABS). Most marketed CSTDs contain either polycarbonate or ABS and are not compatible with Treanda Injection (45 mg/0.5 mL or 180 mg/2 mL solution).

N, N-dimethylacetamide (DMA), an ingredient in Treanda Injection (45 mg/0.5 mL or 180 mg/2 mL solution), is incompatible with polycarbonate or ABS.  Devices that contain polycarbonate or ABS dissolve when coming into contact with DMA. This can lead to device failure, possible product contamination, and potential serious adverse health consequences, including skin reactions in health care professionals preparing and administering this product and the risk of small blood vessel blockage in patients.

FDA is requiring label changes for both the solution and the powder formulations of Treanda to reflect safe preparation information.

BACKGROUND: Treanda is available in two formulations, a solution, Treanda Injection (45 mg/0.5 mL or 180 mg/2 mL solution); and a lyophilized powder, Treanda for Injection (25mg/vial or 100 mg/vial lyophilized powder). Closed system transfer devices are devices that are used to prepare and administer hazardous drugs for intravenous infusion, such as chemotherapy drugs. 

RECOMMENDATION: Health care professionals should stop using Treanda Injection (45 mg/0.5 mL or 180 mg/2 mL solution) with CSTDs or vial adapters and syringes containing polycarbonate or ABS. If using Treanda Injection (45 mg/0.5 mL or 180 mg/2 mL solution):

If using Treanda for Injection (25mg/vial or 100 mg/vial lyophilized powder):

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

Read the MedWatch Safety Alert, including links to the FDA Statement at: http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm437626.htm


This email was sent to list-fda@xxxxxxxxxxx using GovDelivery, on behalf of: U.S. Food & Drug Administration (FDA) · 10903 New Hampshire Ave · Silver Spring, MD 20993 · 800-439-1420 Powered by GovDelivery

[Index of Archives]     [CDC News]     [NIH News]     [USDA News]     [Steve's Art]     [Camping in Yosemite]     [PhotoForum]     [SB Lupus]     [STB]

  Powered by Linux