AngioScore Incorporated Announces Worldwide Voluntary Field Removal of the AngioSculpt® PTA Scoring Balloon Catheter, 0.018" OTW, balloon sizes: 4.0x20mm, 5.0x20mm, and 6.0x20mm

[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

 



Title: AngioScore Incorporated Announces Worldwide Voluntary Field Removal of the AngioSculpt® PTA Scoring Balloon Catheter, 0.018" OTW, balloon sizes: 4.0x20mm, 5.0x20mm, and 6.0x20mm

AngioScore Incorporated Announces Worldwide Voluntary Field Removal of the AngioSculpt® PTA Scoring Balloon Catheter, 0.018" OTW, balloon sizes: 4.0x20mm, 5.0x20mm, and 6.0x20mm
Sun, 09 Jan 2011 09:36:00 -0600

On November 15, 2010 AngioScore Incorporated, Fremont, CA initiated a worldwide recall of the AngioSculpt® PTA Scoring Balloon Catheter, 0.018” OTW, balloon sizes: 4.0x20mm, 5.0x20mm, and 6.0x20mm. This recall was initiated because of a small number of reports from the field of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element, wherein the proximal end of the scoring element remains secured to the catheter.


Manage your FDA Subscriptions:

U.S. Food & Drug Administration (FDA) · 10903 New Hampshire Ave · Silver Spring, MD 20993 · 800-439-1420


[Index of Archives]     [CDC News]     [NIH News]     [USDA News]     [Steve's Art]     [Camping in Yosemite]     [PhotoForum]     [SB Lupus]     [STB]

  Powered by Linux