Abbott Issues Voluntary Recall of POWERSAIL® Coronary Dilatation Catheters

[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

 



Title: Abbott Issues Voluntary Recall of POWERSAIL® Coronary Dilatation Catheters

Abbott Issues Voluntary Recall of POWERSAIL® Coronary Dilatation Catheters
Thu, 30 Jul 2009 10:16:00 -0500

Abbott has conducted a voluntary recall of three lots of POWERSAIL® Coronary Dilatation Catheters from United States distribution and one lot from international distribution as a result of four complaints (one from each lot) that the distal shaft of the catheter exhibited damage. While the issue could be detected and avoided during the preparation for use of the product, it may cause a leak of contrast material during use, which could lead to catheter functional failures and clinical consequences, including air embolism and myocardial infarction, which has the potential to lead to death.

 

Firm Press Release:  FDA posts press releases and other notices of recalls and market withdrawals from the firms involved as a service to consumers, the media, and other interested parties. FDA does not endorse either the product or the company.


Manage your FDA Subscriptions:

GovDelivery, Inc. sending on behalf of U.S. Food & Drug Administration (FDA) · 5600 Fishers Lane · Rockville MD 20857 · 800-439-1420


[Index of Archives]     [CDC News]     [NIH News]     [USDA News]     [Steve's Art]     [Camping in Yosemite]     [PhotoForum]     [SB Lupus]     [STB]

  Powered by Linux